Before an endovascular aneurysm repair, the CT is measured to choose a stent graft that will seal: above the aneurysm in the healthy neck of the aorta, and below it in each common iliac artery. Five measurements decide it. This guide goes through them in the order they're taken, the way the practice grades them.
Scroll down from above the kidneys. The renal arteries leave the aorta to either side, and the neck starts just below the lowest of them, which can be on either side. In the practice, mark that slice as the lowest renal artery.
On every axial CT the patient's right is on the left of the image, so the right renal artery heads left.

Just below the lowest renal, put the calipers across the neck from outer wall to outer wall, thrombus included. Calcium and thrombus sit inside the wall, so even when the bright lumen is narrower, the outer edge is the one to measure to.
If the neck cuts as an oval, it's angled, and the slice is cutting through it on a slant. Measure across the short axis: that's the vessel's real diameter, and the long axis only shows how steep the slant is.

Mark the lowest renal, then the first slice where the aorta widens by more than 10%. That's where the aneurysm starts, and the neck runs between the two. Most devices need at least 10 to 15 mm of it to seal.
The practice measures that length along the vessel's centreline rather than straight down the slices, so a leaning neck comes out a little longer than its slice count.

Scroll through the aneurysm to its widest slice and measure it outer wall to outer wall, across the short axis if it's oval. Include the thrombus: it's part of the aneurysm, even though the bright lumen inside it is narrower.

Below the aortic bifurcation, measure each common iliac outer wall to outer wall where its limb will seal: clear of the bifurcation, and above where the internal iliac comes off. Each side gets its own number, since they're rarely the same.

Each maker publishes a table that turns a measurement into a device size, and the sizes are deliberately a little bigger than the vessel, usually 10 to 20% for the neck, so the graft presses outward and seals. Pick the main body from the neck's diameter and each limb from its iliac.
On the long straight neck case, the 23 mm neck takes a 28 mm main body on Medtronic's Endurant sheet, and the 14 mm right iliac and 13 mm left iliac each take a 16 mm limb.
Mark the aortic bifurcation and each internal iliac origin too, and the practice grades the lengths: each side has to reach at least 15 mm into its common iliac without covering the internal iliac.

Each number is compared with the reference, within a tolerance set by the level.
Two millimetres is about where experienced readers disagree with each other on a diameter. Stars count the measurements alone; points add the graft and your time. A measurement that doesn't count links back to its step here.

The practice teaches the core measurements on axial slices. Planning a real repair goes further, usually in dedicated software such as 3mensio, TeraRecon Aquarius, Siemens syngo.via or OsiriX, and every device has limits in its instructions for use that the case has to fit.
By the ESVS definition, a neck is hostile when it's shorter than 10 mm, wider than 28 mm, angled more than 60 degrees, conical, or heavily lined with thrombus or calcium.
For practice only. Not for planning a real patient's repair: each device's instructions for use are the source of truth.